Zennium X
The Zennium X is the most highly capable electrochemistry workstation carried by Admiral Instruments. It is comprised of a potentiostat / galvanostat and a frequency response analyzer (FRA) unit with a frequency range up to 12 MHz at ±4 A and at a compliance voltage of up to ±32 V. With a standard input impedance of more than 10 TΩ || 5 pF, this workstation is suitable for a wide range of electrochemical measurements, even moreso than the Zennium Pro.
Key Highlights:
-
32-bit converter for exceptional DC accuracy
-
EIS frequency range: 10 µHz to 12 MHz
-
HDP technology with 24-bit AC resolution for high dynamic accuracy
-
Integrated ADF technology enabling ultra-fast CV up to 10 kV/s
-
Compliance voltage up to ±32 V, software-switchable for low noise and sample protection
-
DC voltage ranges ±5 V and ±15 V, optimized for lithium battery research
-
AC amplitude ranges up to 6 V with 18-bit resolution
-
Modular, expandable architecture with optional bipotentiostat
-
Switchable floating/grounded operation
Expandable Architecture:
Zennium X supports up to four plug-in cards (e.g., TEMP/U, EPC42), allowing control of external potentiostats, boosters, or electronic loads. This modular design enables advanced measurement methods and adapts seamlessly as research requirements evolve.
Available Add-Ons:
-
Compliance voltage up to ±120 V
-
Output power up to ±40 A / ±20 V
-
Input impedance up to 1000 TΩ
-
Input current up to 200 A
-
Four additional channels for true parallel EIS
Each Zennium X system includes Zahner’s Thales XT software, providing access to a full suite of standard and advanced electrochemical techniques within a unified platform.

Frequency Generator & Analyzer | Values |
|---|---|
Frequency range | 10 µHz to 12 MHz |
Accuracy | < 0.0025% |
Resolution | 0.0025%, 10000 steps/decade |
Specifications
Output (potentiostatic) | Values |
|---|---|
Controlled voltage | ±5 V, ±15 V |
Resolution | 2.5 nV (5 V range), 7.5 nV (15 V range) |
Accuracy | ±50 µV ± 5 ppm of reading (5 V range), ±200 µV ± 5 ppm of reading (15 V range) |
Integral Nonlinearity | typ. 1 ppm, max. 2 ppm (5 V range), typ. 3 ppm, max. 8 ppm (15 V range) |
Compliance voltage | ±16 V (Low voltage mode), ±32 V (High voltage mode) |
AC amplitude | 0 to 2 V, 18 bit resolution (5 V range),0 to 6 V, 18 bit resolution (15 V range) |
Bandwidth | 15 MHz @ 33 Ω load |
IR compensation | auto AC impedance technique,range 0 to 10 MΩ, resolution 0.012% |
Small signal rise time | 150 ns to 200 µs in 5 steps (autoselect) |
Slew rate | 15 MV/s |
Phase shift | 10° @ 500 kHz |
General Specifications | Values |
|---|---|
Overall bandwidth | DC - 15 MHz |
ADC resolution | 32 bit |
Harmonic reject | > 60 dB @ ½ full scale |
Potentiostat modes | Potentiostatic, galvanostatic, pseudo-galvanostatic, rest potential, off |
Cell connection | 2-, 3-, 4-terminal kelvin |
Ground reference | Grounded, floating |
Extension slots | 10 |
PC interface | USB 2.0 |
Dimensions | 470 x 160 x 376 mm |
Weight | 13.2 kg |
Accessories | cell cable set, USB cable, power cord, manual, Testbox, THALES XT Software Suite |
Power supply | 100/115/230VAC, 50/60Hz |
Ambient temperature | +10 °C to +30 °C |
Ambient humidity | < 60% without derating |
Output (galvanostatic) | Values |
|---|---|
Controlled current | ±4 A |
Current range | ±1 nA to ±4 A in 12 steps |
Resolution | 32 bit, ±0.2 ppb of FS |
Accuracy | 0.01% of FS ± 0.025% of reading (≥1 µA to 100 mA), 0.05 % of FS ± 0.1% of reading (<1 µA or >100 mA) |
Input | Values |
|---|---|
Maximum input voltage | ±5.5 V (5 V range), ±16 V (15 V range) |
Voltage resolution | 2.5 nV (5 V range), 7.5 nV (15 V range) |
Voltage accuracy | ±20 µV ± 2 ppm of reading (5 V range), ±50 µV ± 5 ppm of reading (15 V range) |
Current resolution | ±5 aA (32 Bit) |
Current accuracy | ± 0.01% of FS (1 µA to 100 mA), ± 0.1% of FS (100 mA to 4 A), ± 0.1% of FS (10 nA to 1 µA) |
Input impedance | > 10 TΩ || ±5 pF (typical, standard), > 1000 TΩ || ±1 pF (typical, using HiZ probe) |
Input leakage current | < |± 200 fA| typ., < |± 2 pA| max. |
Impedance range | 1 mΩ to 10 GΩ / 1% (standard), 100 mΩ to 10 MΩ / 0.1%, 100 mΩ to 100 GΩ / 2% (using HiZ probe), 10 µΩ to 1 GΩ / 2% (Gal), 1 mΩ to 10 MΩ / 0.2% |
Common mode rejection | > 86 dB @ 10 µHz to 100 kHz, > 66 dB @ 100 kHz to 12 MHz |
Input channel phase-tracking accuracy | ±0.05 ° @ 10 µHz to 100 kHz, ±0.125 ° @ 100 kHz to 12 MHz |
Equivalent effective input noise | 10 µV rms / 1 pA rms @ 1 mHz to 10 Hz |
Supported Electrochemistry Experiments:
-
Electrochemical Impedance Spectroscopy
-
EIS Series Measurements vs. Parameters
-
Impedance Network Analysis, Simulation & Fitting
-
Impedance/Parameter, Capacity/Voltage Curves
-
AC Voltammetry
-
Cyclic Voltammetry
-
Linear Sweep Voltammetry
-
Triangular Wave Voltammetry
-
Cyclic Triangular Wave Voltammetry
-
Chronoamperometry With Multiple Triangular Potential Sweeps
-
Stationary Electrode Voltammetry
-
Chronoamperometry With Linear Potential Sweep
-
Tafel Scan
-
Steady State Current/Voltage Curves
-
Multi-Cell Current/Voltage Curves
-
Corrosion / Polarisation Measurement
-
Potentiometry
-
Chronopotentiometry
-
Chronoamperometry
-
Pulse Plating
-
Universal Current/Potential/Time Curves
-
Time Domain Controlled Measurements
-
Tast Polarography
-
Differential Pulse Polarography
-
Differential Pulse Voltammetry
-
Stripping Voltammetry
-
Standard Addition Measurement & Analysis
-
Universal Measurement Data Acquisition & Control
-
Universal Frequency Response Analysis