Zennium XC
The ZENNIUM XC is a high-performance electrochemical workstation providing high-end features like a frequency range up to 5 MHz, a maximum current of ±2 A, a controlled voltage of ±14 V, and an input impedance of 10 TΩ || ±5 pF. The ZENNIUM XC is a perfect workstation for a wide range of electrochemical applications with focus on corrosion, coatings and laminates. The ZENNIUM XC can be easily switched from grounded to floating operation.
The ZENNIUM XC comes bundled with the outstanding THALES XT software package which offers all standard methods and more at a mouse click. This is why the Zennium XC can easily be adapted to very different measurement requirements.

Frequency Generator & Analyzer | Values |
|---|---|
Frequency range | 10 µHz to 5 MHz |
Accuracy | < 0.0025% |
Resolution | 0.0025%, 10000 steps/decade |
Specifications
Output (potentiostatic) | Values | Range |
|---|---|---|
Controlled voltage | ±5 V,±14 V | |
Resolution | 2.5 nV, 7.5 nV | 5 V range, 14 V range |
Accuracy | ±200 µV ± 10 ppm of reading,±600 µV ± 10 ppm of reading | 5 V range,14 V range |
Integral Nonlinearity* | typ. 4 ppm, max. 8 ppm,typ. 12 ppm, max. 24 ppm | 5 V range, 14 V range |
AC amplitude | 0 to 2 V, 18 bit resolution, 0 to 6 V, 18 bit resolution | 5 V range, 14 V range |
Bandwidth | 6 MHz @ 33 Ω load | |
IR compensation | auto AC impedance technique,range 0 to 10 MΩ, resolution 0.012% | |
Small signal rise time | 150 ns to 200 µs in 5 steps | automatic selection |
Slew rate* | 15 MV/s | |
Phase shift* | 10° @ 500 kHz |
General Specifications | Values |
|---|---|
Overall bandwidth | DC - 6 MHz |
ADC resolution | 32 bit |
Harmonic reject* | > 60 dB @ ½ full scale |
Potentiostat modes | potentiostatic, galvanostatic, pseudo-galvanostatic, rest potential, ZRA, off |
Cell connection | 2-, 3-, 4-terminal kelvin |
Ground reference | grounded, floating |
PC interface | USB 2.0 |
Dimensions | 255 x 160 x 385 mm |
Weight | 8 kg |
Accessories | 2 cell cable sets, USB cable, power cord, manual |
Power supply | 100/115/230 VAC, 50/60 Hz |
Ambient temperature | +10 °C to +30 °C |
Ambient humidity | < 60% without derating |
Output (galvanostatic) | Values | Range |
|---|---|---|
Controlled current | ±2 A | |
Current range | ±1 nA to ±2 A in 12 steps | Main |
Resolution | 32 bit, ±0.2 ppb of FS (1ppm @ 1 pA) | |
Accuracy | 0.04 % of FS ± 0.1 % of reading, 0.2 % of FS ± 0.4 % of reading | ≥ 1 µA to 100 mA, < 1 µA or > 100 mA |
Input | Values | Range |
|---|---|---|
Maximum input voltage | ±5 V, ±14 V | 5 V range, 14 V range |
Voltage resolution | ±2.5 nV, ±7.5 nV | 5 V range, 14 V range |
Voltage accuracy | ±100 µV ± 5 ppm of reading, ±300 µV ± 10 ppm of reading | 5 V range, 14 V range |
DC current resolution | ±5 aA (32 Bit) | Main / HiZ Probe |
DC current accuracy | ± 0.08% of reading ± 0.04% of FS,
± 0.8% of reading ± 0.4% of FS,
± 0.8% of reading ± 0.4% of FS,
± 0.5% of reading ± 125 fA | 1 µA ... 100 mA,
100 mA ... 2 A,
10 nA ... 1 µA,
< |1 nA| (HiZ Probe) |
Input impedance | > 10 TΩ || ±5 pF (typical),
> 1000 TΩ || ±1 pF (typical) | Main,
HiZ Probe |
Input leakage current | < |± 200 fA| typ.,
< 2 - 5 pA,
< |± 10 fA| typ. | Main,
HiZ Probe |
Impedance range | 1 mΩ to 10 GΩ / 2%,
100 mΩ to 10 MΩ / 0.2%,
100 mΩ to 100 GΩ / 2%,
10 µΩ to 1 GΩ / 2%,
1 mΩ to 10 MΩ / 0.2% | Main,
HiZ,
Gal |
Common mode rejection | > 86 dB @10 µHz to 100 kHz,
> 66 dB @100 kHz to 5 MHz | |
Input channel phase-tracking accuracy | ±0.05 ° @10 µHz to 100 kHz,
±0.125 ° @100 kHz to 5 MHz | |
Equivalent effective input noise | 1 µV rms / 100 fA rms @ 1 mHz to 10 Hz | Main |
Supported Electrochemistry Experiments:
-
Electrochemical Impedance Spectroscopy
-
EIS Series Measurements vs. Parameters
-
Impedance Network Analysis, Simulation & Fitting
-
Impedance/Parameter, Capacity/Voltage Curves
-
AC Voltammetry
-
Cyclic Voltammetry
-
Linear Sweep Voltammetry
-
Triangular Wave Voltammetry
-
Cyclic Triangular Wave Voltammetry
-
Chronoamperometry With Multiple Triangular Potential Sweeps
-
Stationary Electrode Voltammetry
-
Chronoamperometry With Linear Potential Sweep
-
Tafel Scan
-
Steady State Current/Voltage Curves
-
Multi-Cell Current/Voltage Curves
-
Corrosion / Polarisation Measurement
-
Potentiometry
-
Chronopotentiometry
-
Chronoamperometry
-
Pulse Plating
-
Universal Current/Potential/Time Curves
-
Time Domain Controlled Measurements
-
Tast Polarography
-
Differential Pulse Polarography
-
Differential Pulse Voltammetry
-
Stripping Voltammetry
-
Standard Addition Measurement & Analysis
-
Universal Measurement Data Acquisition & Control
-
Universal Frequency Response Analysis