top of page
Squidstat Prime Potentiostat

A cosy-effective, multi-channel potentiostat designed to support a wide range of electrochemical experiments.

Well suited for research and teaching laboratories that benefit from running multiple measurements.

Adobe Express - file (2)_edited_edited_edited.png
1100-ongoing
Serial Numbers:
2-years
Warranty Period:
PR-2X-CS
Item Number: 
  • Four-channel potentiostat/galvanostat

  • Current range: ±250 mA max

  • Voltage scan range: ±10 V

  • Compatible with Windows, MacOS, and Ubuntu via the Squidstat User Interface software and Application Programming Interface

  • Onboard memory (16 GB) automatically backs up all channel data for future recovery

  • Multiple units can run simultaneously with other Squidstat models

  • Made in USA

Description
made-in-usa-logo-vector.png
20 Business Days ARO
Typical Lead Time:

Linear Sweep Voltammetry (LSV)

Normal Pulse Voltammetry

Open Circuit Potential Monitoring

Rotating Ring Disk Electrode (RRDE)

Square Wave Voltammetry

Staircase Voltammetry ​​

Experiments

Battery Capacity

Charge / Discharge (Constant C / Constant V)

Charge / Discharge (Constant R / Constant P)

Chronoamperometry

Chronopotentiometry

Constant Power

Constant Resistance

Cyclic Voltammetry (CV)

Differential Pulse Voltammetry (DPV)​​
Galvanostatic Intermittent Titration Technique (GITT)

Potentiostatic Intermittent Titration Technique (PITT)

Linear Sweep Amperometry

DC Experiments

Output Parameters

DC Parameters

Additional Parameters

Working Electrode (V)
Working Electrode (mV)
DC Working Electrode (V)
DC Working Electrode (mV)
Counter Electrode (V)
Counter Electrode (mV)
Current (A)
Current (mA)

​DC Current (A)
DC Current (mA)
Working Electrode vs. NHE (V)
Current Density (A/m2)

Current Density (A/cm2)
Power (W)
Cumulative Charge (mAh)
Cumulative Charge (C)

Capacity (mAh)
Energy (Wh)
State of Charge (%)
Depth of Discharge (%)
dQ/dV (mAh/V)
1 / Sqrt(time) (1/√s)
Resistance (Ohm)
No of Cycles

PC Timestamp

UTC Time (s)

Potentiostatic Input/Output
Values
Compliance Voltage
±10 V
Voltage Scan Range
±10 V
Number of Voltage Ranges
1 range, [±10 V]
Applied Voltage Accuracy
0.1% of setpoint, 1 mV min
Applied Voltage Resolution
0.003% (30.52 ppm) of range, 300 μV min
Measured Voltage Accuracy
0.1% of measured, 1 mV min
Measured Voltage Resolution
0.003% (30.52 ppm) of range, 300 μV min
Time Resolution
100 µs
iR Compensation Limits
1.25 V
Specifications
Galvanostatic Input/Output
Values
Current Range
±250 mA
Number of Current Ranges
8 ranges, [25 nA to 250 mA]
Applied Current Accuracy
0.1% of range, 1 nA min
Applied Current Resolution
0.003% (30.52 ppm) of range, 760 fA min
Measured Current Accuracy
0.1% of range, 1 nA min
Measured Current Resolution
0.003% (30.52 ppm) of range, 760 fA min
General Specifications
Values
Channels Per Unit
4
Operating Modes
Pot / Gal / ZRA
Cell Connections
5 electrode
ADC/DAC Resolution
16-bit resolution
Slew Rate (1 kΩ load)
35 kV/s (-1 V to +1V)
Shortest DC data recording interval
1000 µs
Shortest DC data hardware sampling interval
100 µs
Floating/Ground Mode
Both (Switchable)
Input Impedance
10 Tera-Ω (typical)
Input Bias Current
10 pA (typical)
Data Backup Memory
16 GB onboard
Dimension and Weight
24 L x 17 W x 6 H cm, 1 kg
Channel Cable Length
1 m
Computer Interface
1 USB per unit
Temperature Operating Range
0 to 40 ℃
Min/Max Scan Rate
300 µV/s minimum, 25 V/s maximum with 10 mV sampling
Smallest Pulse Width
400 µs
Smallest Potential Step for Sweep Voltammetry
0.003% (30.52 ppm) of range, 300 μV min
Square Wave Voltammetry Max Frequency
800 Hz (-1 V to +1V)
Power Supply Requirements
100 – 240 VAC, 50 – 60 Hz
bottom of page