1100-ongoing
Serial Numbers:
2-years
Warranty Period:
SO-2X-CS
20 Business Days ARO
Typical Lead Time:
Item Number:
-
Single-channel potentiostat/galvanostat
-
Current range: ±100 mA max
-
Voltage scan range: ±10 V
-
Compatible with Windows, MacOS, and Ubuntu via the Squidstat User Interface software and Application Programming Interface
-
Onboard memory (16 GB) automatically backs up all channel data for future recovery
-
Multiple units can run simultaneously with other Squidstat models
-
Made in USA
Description

Linear Sweep Voltammetry (LSV)
Normal Pulse Voltammetry
Open Circuit Potential Monitoring
Rotating Ring Disk Electrode (RRDE)
Square Wave Voltammetry
Staircase Voltammetry
Experiments
Battery Capacity
Charge / Discharge (Constant C / Constant V)
Charge / Discharge (Constant R / Constant P)
Chronoamperometry
Chronopotentiometry
Constant Power
Constant Resistance
Cyclic Voltammetry (CV)
Differential Pulse Voltammetry (DPV)
Galvanostatic Intermittent Titration Technique (GITT)
Potentiostatic Intermittent Titration Technique (PITT)
Linear Sweep Amperometry
DC Experiments
Output Parameters
DC Parameters
Additional Parameters
Working Electrode (V)
Working Electrode (mV)
DC Working Electrode (V)
DC Working Electrode (mV)
Counter Electrode (V)
Counter Electrode (mV)
Current (A)
Current (mA)
DC Current (A)
DC Current (mA)
Working Electrode vs. NHE (V)
Current Density (A/m2)
Current Density (A/cm2)
Power (W)
Cumulative Charge (mAh)
Cumulative Charge (C)
Capacity (mAh)
Energy (Wh)
State of Charge (%)
Depth of Discharge (%)
dQ/dV (mAh/V)
1 / Sqrt(time) (1/√s)
Resistance (Ohm)
No of Cycles
PC Timestamp
UTC Time (s)
Potentiostatic Input/Output | Values |
|---|---|
Compliance Voltage | ±12 V |
Voltage Scan Range | ±10 V |
Number of Voltage Ranges | 1 range, [±10 V] |
Applied Voltage Accuracy | 0.1% of setpoint, 2 mV min |
Applied Voltage Resolution | 0.003% (30.52 ppm) of range, 300 μV min |
Measured Voltage Accuracy | 0.1% of measured, 1 mV min |
Measured Voltage Resolution | 0.003% (30.52 ppm) of range, 300 μV min |
Time Resolution | 10 ns |
iR Compensation Limits | 1.82 V max |
Specifications
Galvanostatic Input/Output | Values |
|---|---|
Current Range | ±100 mA |
Number of Current Ranges | 8 ranges, [10 nA to 100 mA] |
Applied Current Accuracy | 0.1% of range, 1 nA min |
Applied Current Resolution | 0.003% (30.52 ppm) of range, 300 fA min |
Measured Current Accuracy | 0.1% of range, 1 nA min |
Measured Current Resolution | 0.003% (30.52 ppm) of range, 300 fA min |
General Specifications | Values |
|---|---|
Channels Per Unit | 1 |
Operating Modes | Pot / Gal / ZRA |
Cell Connections | 5 electrode |
ADC/DAC Resolution | 16-bit resolution |
Slew Rate (1 kΩ load) | 5 kV/s (-1 V to +1V) |
Shortest DC data recording interval | 200 µs |
Shortest DC data hardware sampling interval | 7.5 µs |
Floating/Ground Mode | Both (Switchable) |
Input Impedance | 10 Tera-Ω (typical) |
Input Bias Current | 10 pA (typical) |
Data Backup Memory | 16 GB onboard |
Dimension and Weight | 24 L x 17 W x 6 H cm, 1 kg |
Channel Cable Length | 1 m |
Computer Interface | 1 USB per unit |
Temperature Operating Range | 0 to 40 ℃ |
Min/Max Scan Rate | 300 µV/s minimum, 25 V/s maximum with 10 mV sampling |
Smallest Pulse Width | 375 µs |
Smallest Potential Step for Sweep Voltammetry | 0.003% (30.52 ppm) of range, 300 μV min |
Square Wave Voltammetry Max Frequency | 1 kHz (-1 V to +1V) |
Power Supply Requirements | 100 – 240 VAC, 50 – 60 Hz |